Chapter 5 – Detection and Imaging Tools that Use Nonoptical Waves 167
from the sample. This low energy manifests as a small mean collision path in the sample of
only a few nanometers, and so any secondary electrons that are detected ultimately originate
very close from the sample surface. Thus, SEM using secondary electron detection generates
just a topographical detail of the sample.
Such surface secondary electrons are first accelerated toward an electrically biased grid
at ~90° to the electron beam by a few hundred volts and then further toward a phosphor
scintillator inside a Faraday cage (also known as a Everhart–Thornley detector), coupled to a
photomultiplier tube (PMT) with a higher E-field of ~2 kV potential difference to energize
the electrons sufficiently to allow scintillation in the phosphor. The resulting PMT electric
current is then used as a metric for the secondary electron intensity. Although SEM in itself is
not a 3D technique, the same stage tilting and image reconstruction technology as for trans
mission ET can be applied to generate 3D information on topographical features.
Rarer elastically backscattered electrons are higher in energy and so can scatter at relatively
high angles. The electrons can emerge from anywhere in the sample, and thus, backscattered
electron detection is not a topographic determination technique. To detect backscattered
electrons and not secondary electrons, similar scintillation PMT detectors can be placed in a
ring around the main electron beam (i.e., at relatively high scatter angles), allowing electron
backscatter diffraction images to be generated.
The extent of backscatter is dependent on the atomic number of the metal element in the
contrast reagent. In principle, this offers the potential to apply differential imaging on the
basis of different atomic number components used to stain the sample. This has been applied
to a few exceptional multiple length scale investigations, for example, to probe the optic
nerve tract by using a nonspecific lead metal stain, which reveals topographic information
of the tract from the detected secondary electrons, while using a specific silver metal stain,
which targets just the nerve fibers themselves inside the tract. Silver has a higher atomic
number than lead and thus backscatter electron detection can be used to image just the local
ization of the nerve fibers in the same optic nerve tract.
An SEM can, in principle, be modified to operate simultaneously in the transmission
mode. This involves implementing detectors below the sample to capture transmitted
electrons, as for conventional TEM. Most mainstream EM machines do not operate in this
hybrid manner; however, there is a benefit in using transmission scanning electron micros
copy since, if used in conjunction with LVEM on unstained samples, it improves the image
contrast. Thus, this may serve as a useful control at least against the presence of experimental
artifacts caused through chemical staining procedures.
Some SEM machines are also equipped with an x-ray spectrometer. X-ray spectroscopy
is discussed in more detail later in this chapter, but in essence, K-shell electron ejection also
generates x-rays and their wavelength is dependent on the specific electronic energy levels of
the atom involved. It can therefore be used to investigate the elemental makeup of the sample
(elemental analysis).
Conventional SEM uses the same high vacuum as TEM. The requirement for dehydrated
or frozen samples means that imaging cannot be done under normal “environmental”
conditions. However, the environmental scanning electron microscope (ESEM) overcomes
this limitation to a large extent. ESEM utilizes the same generic SEM design but implements
a modified sample chamber, which allows a higher pressure to be maintained in a humidified
environment. The electron beam attenuation in air increases exponentially with the distance
as the electron beam must penetrate into the sample; therefore, the key developments in
ESEM have been in miniaturization of the sample chamber. Modern ESEM devices often
have variable pressure options with Peltier temperature control for the sample chamber,
allowing a range of EM modes to be used, with pressures of a few kilopascals being sufficient
to prevent water vaporization from wet samples.
5.2.6 CRYO-EM AND CRYOET
The term cryo-EM is often misused in any EM performed on samples, which have been
prepared using cryofixation. However, a better use is for describing EM on a native sample